AFE58JD32ZBV
텍사스 인스트루먼트(주)
- Lifecycle
- Active
- Stock
- 9893 PCS
- Series
- 데이터 수집 - 아날로그 프런트엔드(AFE)
AFE58JD32ZBV과 AFE5808AZCF에 대한 심층 비교를 통해 두 제품의 사양 및 주요 특징에 대한 유용한 정보를 얻을 수 있습니다. RoHS 준수 여부, REACH 인증 상태, 시리즈, 장착 방식, 패키지 유형 및 기타 관련 특성을 포함한 중요한 요소들을 자세히 다룹니다. 두 제품의 차이점을 나란히 제시함으로써 부품 선택이 간소화되어 특정 용도에 가장 적합한 옵션을 쉽게 선택할 수 있습니다.
Replacement verdict
The AFE5808AZCF and AFE58JD32ZBV are not drop-in replacements for one another. They differ in channel count, package footprint, and digital interface architecture, and the AFE58JD32ZBV integrates additional digital processing and JESD204B serial output that the AFE5808AZCF does not provide. A direct board-level substitution is not feasible without redesign of the interface, power, and layout.
텍사스 인스트루먼트(주)
텍사스 인스트루먼트(주)
Rows are prioritized by design impact. Highlighted values require attention during substitution.
Key electrical and mechanical differences between the two devices
| Parameter | AFE5808AZCF | AFE58JD32ZBV | Why it matters |
|---|---|---|---|
| Number of Channels | 8 | 32 | Channel count determines how many transducer elements can be processed simultaneously, directly affecting system architecture and imaging capability. |
| Analog-to-Digital Converter Resolution | 12 bit | 12 bit | ADC resolution affects dynamic range and image quality; both share 12-bit conversion, so this is not a differentiator. |
| Maximum Sampling Rate | 65 MSPS | 65 MSPS | Sampling rate sets the maximum signal bandwidth and axial resolution achievable in the imaging system. |
| Input Referred Noise | 0.75 nV/√Hz | 0.75 nV/√Hz | Low input-referred noise improves sensitivity for weak echo signals, critical for deep-tissue imaging. |
| Total Power Consumption | 158 mW/channel | 90 mW/channel | Per-channel power affects thermal design and portability; the lower-power device enables denser or battery-operated systems. |
| Integrated Continuous Wave (CW) Doppler | Yes | Yes | On-chip CW Doppler supports blood-flow measurement without external circuitry, simplifying system design. |
| Digital Demodulation / Decimation | No | Yes | Integrated digital demodulation reduces FPGA/processor load and data rate, affecting backend complexity and latency. |
| Package | NFBGA-289 | NFBGA-289 | Package footprint and pinout determine PCB layout compatibility and whether a drop-in replacement is feasible. |
| Operating Temperature Range | 0°C to 70°C | 0°C to 70°C | Temperature range defines the environments in which the device can reliably operate. |
Send your BOM and application constraints for technical and sourcing review.
Use manufacturer datasheets as the final authority.
| Specification | AFE58JD32ZBV | AFE5808AZCF |
|---|---|---|
| Part Status | Active | Active |
| Number of Bits | 12 | 12, 14 |
| Number of Channels | 32 | 8 |
| Voltage - Supply, Analog | 3V ~ 3.3V | 1.7V ~ 1.9V, 3.15V ~ 3.6V, 5V |
| Voltage - Supply, Digital | 3V ~ 3.3V | 1.7V ~ 1.9V |
| Mounting Type | Surface Mount | Surface Mount |
| Base Product Number | AFE58JD32 | - |
| Power (Watts) | 42 mW | 158 mW |
The AFE5808AZCF is a Texas Instruments 8-channel analog front-end (AFE) for ultrasound systems with a 14-bit ADC and CW Doppler mixer, supplied in a 289-ball NFBGA (ZCF) package. The AFE58JD32ZBV is a Texas Instruments 32-channel ultrasound AFE with a 12-bit ADC, digital demodulation, and JESD204B serial output, supplied in a 289-ball NFBGA (ZBV) package.
The AFE5808AZCF integrates 8 receive channels, while the AFE58JD32ZBV integrates 32 receive channels, making the AFE58JD32ZBV suited to higher-channel-count ultrasound system architectures.
The AFE5808AZCF uses a 14-bit analog-to-digital converter per channel, whereas the AFE58JD32ZBV uses a 12-bit ADC per channel, so the AFE5808AZCF offers higher amplitude resolution while the AFE58JD32ZBV offers higher channel density.
No. The AFE5808AZCF provides a parallel LVDS output interface, while the AFE58JD32ZBV provides a JESD204B serial output interface, so the digital back-end and FPGA interface requirements differ between the two devices.
No. Although both devices are supplied in a 289-ball NFBGA package, they have different channel counts, ADC resolutions, and output interfaces, so they are not pin-to-pin or functionally interchangeable and require separate PCB designs.
The AFE58JD32ZBV includes digital demodulation and JESD204B output for continuous-wave (CW) Doppler processing, while the AFE5808AZCF provides an analog CW Doppler mixer path; designers should confirm the exact CW Doppler implementation in the respective datasheets.
Both devices target medical ultrasound receive front ends. The AFE5808AZCF is used in 8-channel-class ultrasound systems, while the AFE58JD32ZBV is used in higher-channel-count systems such as 32-channel or larger ultrasound platforms requiring JESD204B serial data links.