AFE58JD28ZAV
텍사스 인스트루먼트(주)
- Lifecycle
- Active
- Stock
- 7063 PCS
- Package
- NFBGA-289
- Series
- 데이터 수집 - 아날로그 프런트엔드(AFE)
AFE58JD28ZAV과 AFE58JD16ZAV에 대한 심층 비교를 통해 두 제품의 사양 및 주요 특징에 대한 유용한 정보를 얻을 수 있습니다. RoHS 준수 여부, REACH 인증 상태, 시리즈, 장착 방식, 패키지 유형 및 기타 관련 특성을 포함한 중요한 요소들을 자세히 다룹니다. 두 제품의 차이점을 나란히 제시함으로써 부품 선택이 간소화되어 특정 용도에 가장 적합한 옵션을 쉽게 선택할 수 있습니다.
Replacement verdict
The AFE58JD28ZAV and AFE58JD16ZAV share the same Texas Instruments ultrasound analog front-end family, package footprint, and supply architecture, making them viable functional replacement candidates. However, the AFE58JD28ZAV offers twice the channel count and higher per-channel power, so system-level validation of channel mapping, power budget, and thermal performance is required before substitution.
텍사스 인스트루먼트(주)
텍사스 인스트루먼트(주)
Rows are prioritized by design impact. Highlighted values require attention during substitution.
Key electrical and mechanical differences between the two devices
| Parameter | AFE58JD28ZAV | AFE58JD16ZAV | Why it matters |
|---|---|---|---|
| Number of receive channels | 32 | 16 | Channel count determines how many transducer elements can be digitized simultaneously, directly affecting imaging system architecture and the number of AFEs required per system. |
| ADC resolution | 14 bit | 12 bit | Higher ADC resolution provides greater dynamic range and finer quantization, improving image quality and small-signal detection in ultrasound receive chains. |
| Maximum ADC sampling rate | 65 MSPS | 65 MSPS | Sampling rate sets the maximum signal bandwidth that can be digitized, which constrains the achievable imaging depth and axial resolution. |
| LVDS serial data rate per channel | Up to 1 Gbps | Up to 1 Gbps | The serial output rate determines the digital interface bandwidth needed to stream digitized channel data to the downstream beamformer or FPGA. |
| Integrated CW Doppler beamformer | Yes (with 16-bit CW mixer and decimation filter) | Yes (with 16-bit CW mixer and decimation filter) | On-chip continuous-wave Doppler processing offloads the FPGA and simplifies the design of systems requiring blood-flow velocity measurement. |
| Digital time-gain compensation (TGC) range | Up to 48 dB | Up to 48 dB | TGC compensates for depth-dependent signal attenuation; the available range affects the ability to equalize echoes from near and far field targets. |
| Programmable gain amplifier (PGA) range | 5 dB to 40 dB | 5 dB to 40 dB | PGA gain range sets the input signal window that can be amplified before the ADC, influencing sensitivity and overload margin. |
| Package | 289-ball NFBGA (ZAV) | 289-ball NFBGA (ZAV) | Package and pinout compatibility determine whether the two devices can share a common PCB footprint or require a board redesign. |
| Supply voltage | 1.8 V / 3.3 V (analog and digital rails) | 1.8 V / 3.3 V (analog and digital rails) | Supply rail requirements affect power supply design and whether the parts can be interchanged without changing the power architecture. |
| Operating temperature range | -40°C to +85°C | -40°C to +85°C | The temperature range confirms suitability for medical and industrial environments and must match the target application's environmental requirements. |
Send your BOM and application constraints for technical and sourcing review.
Use manufacturer datasheets as the final authority.
| Specification | AFE58JD28ZAV | AFE58JD16ZAV |
|---|---|---|
| Supplier | - | Rochester Electronics, LLC,Texas Instruments |
| Part Status | Active | Active |
| Number of Bits | 12, 14 | 14 |
| Number of Channels | 16 | 16 |
| Power (Watts) | 64 mW | 90 mW |
| Mounting Type | Surface Mount | Surface Mount |
| Voltage - Supply, Analog | 3.15V ~ 3.6V | - |
Both devices are 16-channel ultrasound AFEs in the same ZAV package family, but the AFE58JD28ZAV provides 14-bit ADC resolution while the AFE58JD16ZAV provides 12-bit ADC resolution. The AFE58JD28ZAV also offers a higher maximum sampling rate (65 MSPS vs 50 MSPS for the AFE58JD16ZAV) and lower minimum power per channel, making it the higher-performance option for demanding imaging systems.
Both devices are offered in the same ZAV package (NFBGA) with a 16-channel signal chain architecture, and they share the same functional block arrangement (LNA, VCA, PGA, ADC, CW mixer, and digital demodulation). However, designers must confirm pin function assignments and power supply requirements against the respective TI datasheets before treating them as drop-in replacements, since ADC resolution and timing specifications differ.
The AFE58JD28ZAV integrates a 14-bit ADC with a maximum sampling rate of 65 MSPS. The AFE58JD16ZAV integrates a 12-bit ADC with a maximum sampling rate of 50 MSPS. These differences directly affect dynamic range and image quality in ultrasound receive chains.
Yes. Both the AFE58JD28ZAV and AFE58JD16ZAV integrate an on-chip continuous wave (CW) Doppler mixer with a low-power mode, supporting CW processing for Doppler flow measurements without external mixing components.
Both devices include digital I/Q demodulation, decimation filters, and a low-voltage differential signaling (LVDS) output interface for transferring digitized data to an FPGA or DSP. The AFE58JD28ZAV supports higher output data rates consistent with its 65 MSPS maximum sampling rate, while the AFE58JD16ZAV supports rates up to its 50 MSPS limit.
The AFE58JD28ZAV is specified at approximately 88 mW per channel in continuous wave mode and around 105 mW per channel in normal imaging mode. The AFE58JD16ZAV is specified at approximately 100 mW per channel in normal imaging mode. Designers should refer to the latest TI datasheets for exact power figures under specific operating conditions, as these values depend on sampling rate and mode configuration.
The AFE58JD28ZAV is generally preferable for high-performance portable systems because its 14-bit resolution and 65 MSPS sampling rate deliver better dynamic range and image quality while maintaining competitive power per channel. The AFE58JD16ZAV may be selected where 12-bit resolution and 50 MSPS are sufficient and cost or power optimization favors the lower-tier device.
Both devices feature a 16-channel receive chain consisting of a low-noise amplifier (LNA), a voltage-controlled attenuator (VCA), a programmable gain amplifier (PGA), and an analog-to-digital converter (ADC), followed by digital demodulation and decimation. This common architecture allows similar system-level design approaches, with performance differences driven primarily by ADC resolution and maximum sampling rate.